Model: MB120
Maximum Capacity: 120 g
Moisture Content Readability: 1 mg / 0.01%
Pan Size: 3.5 in (90 mm)
Item Number: 30241165
Overview
Moisture content significantly impacts the quality and functionality of everyday materials. Precise moisture measurement is crucial in various applications, and the MB120 moisture analyzer is designed to make this process faster, smarter, and more efficient. This advanced instrument handles both routine and complex measurements, allowing you to analyze more samples, easily access stored data, and focus more on utilizing your results.
Applications
- Moisture Content Determination
- Moisture Regain Determination
- Solid Content Determination
- General Weighing
Display
- 4.3″ TFT color touchscreen for intuitive operation
Operation
- Powered by an included power cord
Communication
- Equipped with RS232 for GLP/GMP data output
- USB Host and USB Device ports included
Construction
- Sturdy metal base with ABS top housing
- Halogen heat source
- Stainless steel pan support and handler
- Protective in-use cover
Design Features
- Seven shut-off options: manual, timed, and five automatic modes based on weight or moisture loss over time
- Four drying profiles: standard, fast, ramp, and step
- Storage capacity for up to 100 methods and 1,000 test results
- Heating range from 40°C to 230°C, adjustable in 1°C increments
- Available in 14 languages for versatile use
Ohaus MB120 Moisture Analyzer Specifications
- Maximum Capacity: 120 g
- Moisture Content Readability: 1 mg / 0.01%
- Pan Size: 3.5 in (90 mm)
- Heater Type: Halogen Lamp
- Communication: USB device, USB host x 2, RS232
- Dimensions: 13.98 in x 7.09 in x 8.31 in (355 mm x 180 mm x 211 mm)
- Display: 4.3 in Full-Color Touchscreen WQVGA
- Heating Profiles: Ramp, Standard, Step, Fast
- In-Use Cover: Included
- Moisture Range: 0% – 100%
- Net Weight: 11.53 lb (5.23 kg)
- Operating Temperature Range: 10°C – 40°C
- Power Supply: 115 V, 60 Hz
- Sample Size: 0.5 g – 120 g
- Shut-Off Criteria: Timed, Manual, or Customizable based on weight or moisture loss over time
- Temperature Range: 40°C – 230°C
- Test Library: Capable of storing 100 methods
- Units of Measurement: %DC, Gram, %RG, %MC
- Working Environment: 50°F – 104°F (10°C – 40°C), up to 85% RH, non-condensing