CO2 Incubator ICO50 Memmert – Technical Data
Temperature
- Working-temperature range: 5 °C above ambient temperature up to +50 °C
- Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
- Setting temperature range: +18 to +50 °C
- Setting accuracy temperature: 0.1 °C
- Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
- Temperature variation in chamber at +37 °C: +/- 0.3 K
- Temperature fluctuations with time at 37 °C: +/- 0.1 K
Humidity
- Humidity control (standard): Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
- Setting accuracy humidity: 0.5 % rh
- Setting range active humidity control (with option K7): 40 to 97 % rh and rh-Off
Control of Standard Components
- CO2 control: Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
- Adjustment range CO2: 0 to 20 % CO2
- Setting accuracy CO2: 0.1%
- Variation in time CO2: +/- 0.2 % CO2
- Adjustment range O2: 1 to 20 % O2
- Setting accuracy O2: 0.1 % O2
Control Technology
- ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
- Language setting: German, English, Spanish, French, Polish, Czech, Hungarian
- Function SetpointWAIT: The process time does not start until the set temperature is reached
- Adjustable parameters: Temperature (Celsius or Fahrenheit), CO2, programme time, time zones, summertime/wintertime
Communication
- Interface: Ethernet LAN, USB
- Documentation: Programme stored in case of power failure
- Programming: AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Safety
- AutoSAFETY: Additionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
- Autodiagnostic system: Integral fault diagnostics for temperature and CO2
- Alarm: Visual and acoustic
Heating Concept
- 6 sides: Large-area multi-function heating system on four sides with additional door and back heating to avoid condensation
Standard Equipment
- Works calibration certificate:
- At +37 °C, 5 % CO2 and 90 % rh (requires option K7); standard equipment for units with active humidity control
- At +37 °C, 5 % CO2, 90 % rh and 10 % O2 (requires option K7 and option T6); standard equipment for units with O2 control
- Incl. works calibration certificate (measuring point chamber centre) at +37°C, 5 % CO2 for standard units
- Door: Fully insulated stainless steel door with 2-point locking (compression door lock)
- Inner door: Inner glass door with opening (8 mm Ø) to take gas sample
- CO2 connection set: Hose with coupling and clamp
- Standard accessories: Membrane filter (in order to remove impurities and pollutants, all incoming gases pass through a membrane filter before they reach the chamber)
- Internals: 1 perforated stainless steel shelf/shelves
- Water dish: 1 stainless steel water dish
Stainless Steel Interior
- Dimensions (w x h x d): 400 x 425 x 330 mm (d less 35 mm for fan)
- Material: 1.4301 (ASTM 304), corrosion resistant
- Volume: 56 l
- Max. number of internals: 5
- Max. loading of chamber: 75 kg
- Max. loading per internal: 15 kg
Textured Stainless Steel Casing
- Dimensions (w x h x d): 559 x 795 x 521 mm (d +56 mm door handle)
- Housing: Rear zinc-plated steel
Electrical Data
- Voltage Electrical load:
- 230 V, 50/60 Hz, approx. 1100 W
- 115 V, 50/60 Hz, approx. 1100 W
Ambient Conditions
- Set Up:
- The distance between the wall and the rear of the appliance must be at least 15 cm.
- The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
- Ambient temperature: 10 °C to 35 °C
- Humidity rh: Max. 70 %, non-condensing
- Altitude of installation: Max. 2,000 m above sea level
- Overvoltage category: II
- Pollution degree: 2
Packing/Shipping Data
- Transport information: The appliances must be transported upright
- Customs tariff number: 8419 8998
- Country of origin: Federal Republic of Germany
- WEEE-Reg.-No.: DE 66812464
- Dimensions approx incl. carton (w x h x d): 730 x 950 x 640 mm
- Net weight: Approx. 55 kg
- Gross weight carton: Approx. 74 kg