Lianhua Biochemical oxygen demand meter LH-BOD606
The Lianhua Biochemical oxygen demand meter LH-BOD606 measures biochemical oxygen demand in water and wastewater samples for professional laboratory testing. This BOD5 analyzer supports 6 independent test groups, a measurement range of 0–4000 mg/L, and a flexible measurement period from 1 to 30 days. The LHOS operating system, wireless communication, 16G SD card storage, and reliable pressure measurement performance help laboratories manage BOD testing more efficiently. The LH-BOD606 works well for water quality testing, wastewater treatment monitoring, environmental analysis, and laboratory research.
Features
- The LH-BOD606 measures biochemical oxygen demand in water quality analysis applications.
- The instrument follows the national standard HJ505-2009 reaction process and ISO 9408-1999 manufacturing basis.
- The 0–4000 mg/L measurement range supports various water and wastewater samples.
- The analyzer runs 6 independent groups of tests, so users can measure multiple samples at the same time.
- Users can set the measurement period from 1 to 30 days for BOD5 testing and extended BOD analysis.
- The instrument records results every 1 hour to help users monitor test progress during the measurement cycle.
- The display shows 0.1 mg/L resolution for readings below 10 mg/L and 1 mg/L resolution for readings of 10 mg/L or above.
- The pressure measurement accuracy of ≤±2.5% and measurement accuracy of ≤±10% support dependable routine laboratory results.
- The air tightness of less than 0.1 kPa/15 min helps maintain stable test conditions during BOD analysis.
- The 16G SD card gives users large storage capacity for test records and historical data.
- Wireless communication helps users transfer data more conveniently.
- The culture temperature of 20 ±1°C supports standard BOD testing requirements.
Specifications
- Product Name: Biochemical Oxygen Demand BOD5 Analyzer
- Instrument Model: LH-BOD606
- Standard Basis: Follows the national standard HJ505-2009 reaction process and ISO 9408-1999
- Display Resolution: 0.1 mg/L for readings below 10 mg/L; 1 mg/L for readings of 10 mg/L or above
- Operating System: LHOS operating system
- Measuring Range: 0–4000 mg/L
- Pressure Measurement Accuracy: ≤±2.5%
- Air Tightness: <0.1 kPa/15 min
- Measurement Accuracy: ≤±10%
- Frequency of Recording Results: 1 hour
- Measurement Period: 1–30 days, optional
- Measurement Data: 6 independent groups of tests
- Culture Bottle Capacity: 580 mL
- Data Storage: 16G SD card storage
- Communication Interface: Wireless communication
- Culture Temperature: 20 ±1°C
- Rated Power: 30 W
- Power Configuration: 100–240 V, 50–60 Hz
- Instrument Size: 306 × 326 × 133 mm
- Instrument Weight: 6.3 kg
- Ambient Temperature: 5–40°C
- Environmental Humidity: ≤85% RH
Applications
- Lianhua Biochemical Oxygen Demand Meter LH-BOD606 supports BOD testing for wastewater, surface water, industrial process water, domestic wastewater, river water, lake water, and environmental water samples.
- Environmental laboratories, wastewater treatment plants, industrial quality control departments, water monitoring stations, universities, research institutions, and testing centers can use this BOD5 analyzer for routine analysis.
- The LH-BOD606 helps users evaluate organic pollution levels, monitor wastewater treatment efficiency, support environmental compliance testing, and perform routine water quality analysis.
- Its 6 independent test groups and large data storage capacity help laboratories run BOD measurements more efficiently and organize test records more easily.