The Micro balance AND LCC-11-T010-K is a 10-ton capacity compression load cell engineered for industrial-grade weighing applications such as tank, hopper, and truck scale systems. Designed with a hermetically sealed stainless steel body, this model is fully protected against moisture, dust, and corrosive environments, meeting IP68 standards for continuous operation in wet and harsh outdoor conditions.
Built with an advanced load design, the LCC-11-T010-K ensures excellent linearity, high output stability, and long-term performance under fluctuating environmental and loading conditions.
Key Features
10-Ton Compression Capacity (98.07 kN)
Suitable for applications that require stable, vertical load measurement, such as vessel weighing, silo monitoring, and vehicle scales.High Signal Accuracy
Rated Output: 2 mV/V ± 0.1%
Combined Error: ≤ 0.016% of Rated Output
Zero Balance: ±1% of Rated Output
Ensures high-precision performance with minimal measurement drift.
IP68 Waterproof Protection
Fully sealed stainless steel body is resistant to water immersion, dust ingress, and chemical corrosion, making it ideal for food plants, outdoor use, or washdown zones.Temperature Stability
Operating Range: –20°C to +60°C
Zero Temperature Effect: 0.019% of R.O./10°C
Span Temperature Effect: 0.010% of Load/10°C
Robust Electrical Design
Insulation Resistance: >5000 MΩ @ 50 V DC
Designed to withstand electrically noisy industrial environments
Integrated Heavy-Duty Cable
Comes with a 12-meter, Ø8 mm, 4-core shielded cable, ready for field deployment without additional accessories.
Typical Applications
Truck and Vehicle Scales
A trusted solution for multi-load cell systems used in weighbridges and mobile platforms.Tank, Hopper, and Silo Weighing
Enables precise batch control, filling, and level detection in bulk material management.Outdoor and Harsh Industrial Use
Performs reliably under exposure to water, dust, chemicals, and fluctuating temperatures.Legal-for-Trade Installations (with approved systems)
Suitable for commercial use in regulated weight monitoring and billing systems.
Technical Specifications
| Specification | Value |
|---|---|
| Model | LCC-11-T010-K |
| Rated Capacity | 10 tf (98.07 kN) |
| Rated Output | 2 mV/V ± 0.1% |
| Combined Error | ≤ 0.016% of Rated Output |
| Zero Balance | ±1% of Rated Output |
| Input Resistance | 800 ± 80 Ω |
| Output Resistance | 2200 ± 10 Ω |
| Insulation Resistance | >5000 MΩ @ 50 V DC |
| Recommended Excitation Voltage | 5 to 12 V DC |
| Maximum Excitation Voltage | 15 V DC |
| Maximum Safe Overload | 200% of Rated Capacity |
| Operating Temperature Range | –20°C to +60°C |
| Zero Temp Effect | 0.019% of R.O./10°C (typical) |
| Span Temp Effect | 0.010% of Load/10°C (typical) |
| Environmental Protection | IP68 (hermetically sealed stainless steel) |
| Cable | Ø8 mm, 4-core shielded, 12 m length |
| Weight | ~8.5 kg (est. based on series) |