Standgas Pro LCD TOX Duran Electronica

Rp123

Tanyakan Harga dan Stok Produk Terlebih Dahulu Melalui Whatsapp / Email Kami!

Standgas Pro LCD TOX Duran Electronica

The STANDGAS PRO LCD TOX is a standalone, programmable detector designed for the detection of toxic gases and oxygen. It employs electrochemical technology for toxic gases and infrared technology for detecting CO2 and A1 safety-grade refrigerants, with additional early detection of hydrogen in battery rooms.

This detector features:
  • A 16 x 2 line backlit display
  • Three control keys
  • Programmable alarm relay output
  • Internal acoustic alerts
Special Functions for Toxic Gases:
  • Digital monitoring of sensor status
  • Automatic hardware status checks
  • Digital filtering with variable sampling of the sensor’s average values
  • Automatic temperature compensation for accurate response to temperature changes
  • Infrared sensors with microprocessor control for temperature compensation, linearization, and calibration parameter storage
  • Easy installation replacement without needing recalibration or additional gas
Additional Features:
  • Standalone operation without needing a control center
  • Mixed functionality: local information and operation with RS485 communication to a control center
  • IP65 protection grade
  • Optional universal support for ceiling or floor installation
  • Optional 230V power charger that can connect 1 to 10 STANDGAS detector units
Available Gases:
  • Carbon monoxide (CO)
  • Nitrogen dioxide (NO2)
  • Ammonia (NH3)
  • Hydrogen sulfide (H2S)
  • Chlorine (Cl2)
  • Carbon dioxide (CO2)
  • Sulfur dioxide (SO2)
  • Sulfur hexafluoride (SF6)
  • Oxygen (O2)
  • Hydrogen (H2)
  • Refrigerants: R-507, R-125, R-404a, R-407a, R-407f, R-410a, R-449, R-417a, R-448a, R-227ea, R-1233zd, R-513a, R-422d, R-452a, R-134a (for other gases, please inquire)

Applications:

  • Pharmaceutical laboratories
  • University labs
  • Aircraft industry
  • Galleries
  • Distilleries
  • General refrigeration installations
  • Livestock farming installations
  • Boiler rooms
  • Battery rooms
  • Car parks or electric buses
  • General industry
  • Early detection of hydrogen leaks from lithium-ion batteries

Factory Settings for STANDGAS PRO LCD TOX:

GasInstallation HeightRangePrealarmAlarmStandby RelayResetAcousticCovered Area
Carbon Monoxide (CO)1.50/2m. Floor0-300 ppm50 ppm70 ppmDisabledAutoYesApprox. 200 m²
Nitrogen Dioxide (NO2)40/50cm. Floor0-20 ppm3 ppm5 ppmDisabledAutoYesApprox. 100 m²
Ammonia (NH3)30/40cm. Ceiling0-100 ppm5 ppm8 ppmDisabledAutoYesApprox. 75 m²
Hydrogen Sulfide (H2S)1.5m. Floor0-100 ppm5 ppm10 ppmDisabledAutoYesApprox. 100 m²
Carbon Dioxide (CO2)1m. Floor0-20,000 ppm5,000 ppm10,000 ppmDisabledAutoYesApprox. 100 m²
Oxygen (O2)1.70m. Floor0-25%19%18%DisabledAutoYesApprox. 100 m²
Chlorine (Cl2)1m. Floor0-10 ppm0.5 ppm1 ppmDisabledAutoYesApprox. 100 m²
Sulfur Dioxide (SO2)30/40cm. Floor0-20 ppm2 ppm5 ppmDisabledAutoYesApprox. 75 m²
Hydrogen (H2)30/40cm Ceiling0-500 ppm50 ppm100 ppmDisabledManualYesApprox. 25 m²

Refrigerant Gases:

Refrigerant GasRangePrealarmAlarmStandby RelayResetAcousticCovered Area
Refrigerants and SF60-2,000 ppm300 ppm600 ppmDisabledManualYesNot Specified

 

Standgas Pro LCD TOX Duran Electronica Technical Characteristics

  • Technology: Electrochemical / Infrared Sensor + 12-Bit Microprocessor
  • Supply Voltage:
    • Standard: 10V to 24V DC
    • RS485 Version: 10V to 15V
  • Maximum Consumption at 12V DC:
    • Standby: Approx. 43 mA
    • Acoustic and Relay Activated: Approx. 73 mA
    • CO2: 30 mA
    • RS485 Version: 127 mA Max.
  • Communication: RS485
  • Gas Measurement Range: Refer to Factory Settings Table; Up to 2000 ppm for refrigerants
  • Resolution:
    • Toxic Gases: ±1% F.S.
    • Oxygen (O2): ±0.25%
    • Carbon Dioxide (CO2): ±1.5%
    • Refrigerants: ±1%
  • Repeatability: ±2% F.S. for most gases; ±0.5% for Hydrogen (H2)
  • Annual Span Drift:
    • Electrochemical Gases: ±<1%
    • CO2: ±0.7%
    • Refrigerants: ±0.1% F.S.
  • Stabilization Time: <5 minutes for all specifications
  • T90 Response Time:
    • CO, NO2, SO2, NH3: ≤30s
    • H2S: ≤20s
    • H2: ≤60s
    • O2: ≤15s
    • CO2: ≤15s
    • Refrigerants: ≤30s
  • Useful Life (MTBF):
    • Electrochemical Gases (NH3, H2S, NO2, CI2, SO2, H2, O2): Approx. 2 years
    • CO: Approx. 4 years
    • CO2 and Refrigerants: >5 years
  • Maintenance Periods: Recommended annually or as per current standards
  • Environmental Conditions:
    • Temperature: -10ºC to +50ºC
    • Humidity: 15% to 90% R.H. (without condensation)
  • Atmospheric Pressure Limits: 80 to 110 kPa (0.8 to 1.1 bar)
  • Alarm Relay: Fuse-protected dry contact switch output, 3A 250V AC
  • Coverage Area: Refer to Factory Settings Table
  • Material and Protection Grade: Makrolon & ABS, IP65
  • Input and Cable Diameter: Cable glands / 6-10 mm²
  • Dimensions (mm): 120 x 160 x 60
  • Weight (g): 350

Reviews

There are no reviews yet.

Be the first to review “Standgas Pro LCD TOX Duran Electronica”